Por favor, use este identificador para citar o enlazar este ítem: http://ipicyt.repositorioinstitucional.mx/jspui/handle/1010/1973
Removing lateral chromatic aberration in bright field optical microscopy
MIGUEL ANGEL GUZMAN ALTAMIRANO
Braulio Gutiérrez Medina
Acceso Abierto
Atribución-NoComercial-SinDerivadas
https://doi.org/10.1364/OE.23.014380
Fluorescence microscopu
"We present an efficient alternative to remove lateral chromatic aberration (LCA) in bright field light microscopy images. Our procedure is based on error calibration using time-sequential acquisition at different wavelengths, and error correction through digital image warping. Measurement of the displacements of fiducial marks in the red and green images relative to blue provide calibration factors that are subsequently used in test images to realign color channels digitally. We demonstrate quantitative improvement in the position and boundaries of objects in target slides and in the color content and morphology of specimens in stained biological samples. Our results show a reduction of LCA content below the 0.1% level. "
Optical Society
2015
Artículo
Miguel Guzmán-Altamirano and Braulio Gutiérrez-Medina, "Removing lateral chromatic aberration in bright field optical microscopy," Opt. Express 23, 14380-14390 (2015)
OPTICA FÍSICA
Versión publicada
publishedVersion - Versión publicada
Aparece en las colecciones: Publicaciones Científicas Nanociencias y Materiales

Cargar archivos:


Fichero Tamaño Formato  
OpticsExpress23(2015)14380.pdf2.3 MBAdobe PDFVisualizar/Abrir